Department of Pharmaceutical Analysis and Quality Assurance, Bapatla College of Pharmacy, Bapatla-522101, Andhra Pradesh, India.
*E-mail: marellavijayalakshmi@yahoo.co.in
https://doi.org/10.53879/id.49.07.p0054
ABSTRACT
A RP-HPLC method was developed and validated for the determination of ramosetron hydrochloride in pharmaceutical formulation as per ICH and FDA guidelines. The method was carried out on a Phenomenex RP-C18 column using a mixture of methanol and water (95:5) in an isocratic mode. The flow rate is 0.8 mL/min and the detection was done at 302 nm. The linearity range was observed in the range of 1-6 mcg/mL. The accuracy of the method was found to be 99.0 to 99.5% and %RSD was found to be less than 2% indicating high degree of accuracy and precision for the proposed HPLC method. Limit of detection and limit of quantification of the method were found to be 0.028 and 0.0851 mcg/mL respectively.